I-GDOT-80A Insulation Oil Tester manual-updated1105
●Sicela ufunde imanuwali yokusebenza ngokucophelela ngaphambi kokusebenza.
●Sicela uhlole ukuthi umhloli uxhumeke kahle yini emhlabeni ngaphambi kokuhlolwa.
●Akuvunyelwe ukuhambisa noma ukuphakamisa ikhava yokuhlola ngesikhathi sokuhlolwa ukuze kugwenywe ukulimala kwamandla kagesi aphezulu.Amandla kufanele acinywe ngaphambi kokufaka uwoyela wesampula.
●Bamba ngokucophelela lapho usuka noma uvala ikhava yokuhlola amandla kagesi aphezulu!
●Uma isihloli sisebenza ngendlela engavamile ngemva kokuphuka kukawoyela we-insulating, sicela ucime isihloli imizuzwana eyi-10, bese uqala kabusha futhi.
●Ngemuva kokuthi iphepha lokuphrinta seliphelile, sicela ubheke ingxenye yencazelo yephrinta (noma isithasiselo semanuwali) ukuze ungene esikhundleni sephepha lokuphrinta ukugwema ukulimala kwekhanda lephrinta.
●Gcina isihloli sikude nomswakama, uthuli nezinye izinto ezilimazayo, futhi usigcine kude nemithombo yokushisa ephezulu.
●Phatha ngokucophelela kwezokuthutha.Ungabeki ubheke phansi.
●Imanuwali ingabuyekeza ngokufanele ngaphandle kwesinye isaziso kusenesikhathi.Uma kukhona umbuzo, sicela usithinte.
Isikhathi sewaranti salolu chungechunge unyaka owodwa kusukela ngedethi yokuthunyelwa.Sicela ubhekisele ku-invoyisi yakho noma amadokhumenti okuthumela ukuze unqume izinsuku ezifanele zesiqinisekiso.I-HVHIPOT corporation iqinisekisa kumthengi wangempela ukuthi lo mkhiqizo ngeke ungabi nasici ezintweni ezibonakalayo kanye nomsebenzi owenziwayo ngaphansi kokusetshenziswa okuvamile.Ngaso sonke isikhathi sewaranti, hlinzeka ngokuthi lokho kulimala akunqunywanga i-HVHIPOT ukuthi kudalwe ukuhlukumeza, ukusetshenziswa kabi, ukuguqulwa, ukufakwa okungafanele, ukunganaki noma isimo esibi semvelo, i-HVHIPOT inomkhawulo kuphela wokukhanda noma ukushintshwa kwaleli thuluzi phakathi nesikhathi sesiqinisekiso.
Insimbi ye-GDOT-80C | 1 pc |
Inkomishi yamafutha (250ml) | 1 ama-pcs |
Intambo yamandla | 1 pc |
Gcina i-fuse | 2 ama-pcs |
Induku enyakazayo | 2 ama-pcs |
I-Standard gauge(25mm) | 1 pc |
Iphepha lokuphrinta | 2 imiqulu |
I-Tweezer | 1 pc |
Igayidi yomsebenzisi | 1 pc |
Umbiko wokuhlolwa kwefekthri | 1 pc |
I-HV Hipot Electric Co., Ltd. iye yalihlola ngokucophelela nangokucophelela le manuwali, kodwa ngeke siqinisekise ukuthi awekho amaphutha nokushiywa ngokuphelele encwadini.
I-HV Hipot Electric Co., Ltd. izibophezele ekwenzeni ngcono okuqhubekayo emisebenzini yomkhiqizo, kanye nokwenza ngcono ikhwalithi yesevisi, ngakho inkampani ihlala inelungelo lokushintsha noma yimiphi imikhiqizo nezinhlelo zesofthiwe ezichazwe kuleli bhukwana kanye nokuqukethwe kwaleli bhukwana ngaphandle kwangaphambi kwalokho. isaziso.
Ukufakwa kwangaphakathi kwemishini kagesi eningi ezinhlelweni zamandla, izinhlelo zikaloliwe, izitshalo ezinkulu ze-petrochemical kanye namabhizinisi, kwamukelwa kakhulu uhlobo lokufakelwa olugcwele uwoyela, ngakho-ke, ukuhlolwa kwamandla e-dielectric kawoyela okuvikelayo kujwayelekile futhi kuyadingeka.Ukuze kuhlangatshezwane nezidingo zemakethe, sithuthukise futhi sakhiqiza uchungechunge lwabahloli bamandla e-dielectric kawoyela afaka uwoyela ngokwezinga likazwelonke le-GB/T507-2002, izinga lemboni i-DL429.9-91 kanye ne-Electric Power Industry Standard DL/T846.7 yakamuva -2004 ngokwethu.Leli thuluzi, ngokusebenzisa i-single-chip microcomputer njengomgogodla, laqaphela ukusebenza okuzenzakalelayo, ukulinganisa okunembile okuphezulu, lithuthukisa kakhulu ukusebenza kahle komsebenzi kanye nokunciphisa umfutho wokusebenza kwabasebenzi.Ngaphezu kwalokho, incane ngosayizi futhi kulula ukuyithwala.
●Nge-microprocessor, gcwalisa ngokuzenzakalelayo ukuhlolwa kwe-voltage yokumelana nokujikeleza kukawoyela ngebanga elingu-0~80KV (okuhlanganisa ukukhulisa, ukunakekela, ukuhlanganisa, ukuma, ukubala, ukuphrinta nokunye ukusebenza).
●Isibonisi esikhulu sesikrini se-LCD.
●Ukusebenza okulula.Umshini uzoqedela ngokuzenzakalelayo ukuhlolwa kokumelana ne-voltage enkomishini eyodwa kawoyela oyisampula ngemva kokusetha okulula okwenziwa ngu-opharetha.Inani le-voltage ehlukanisiwe lezikhathi ezingu-1~6 nezikhathi zomjikelezo zizolondolozwa ngokuzenzakalelayo.Ngemuva kokuhlolwa, iphrinta eshisayo izophrinta inani ngalinye le-voltage ehlukanisiwe kanye nenani elimaphakathi.
●Ukulondolozwa kwamandla phansi.Ingakwazigcina imiphumela yokuhlolwa engu-100 futhi ubonise izinga lokushisa lamanje le-ambient kanye nomswakama.
●Yamukela i-single-chip microcomputer ukuze ukhuphule i-voltage ngesivinini esingaguquki.I-voltage frequency inembile ku-50HZ, qinisekisa ukuthi yonke inqubo kulula ukuyilawula.
●Nge-over-voltage, over-current and limit limit protection ukuqinisekisa ukuphepha kwabaqhubi.
●Ngomsebenzi wokubonisa izinga lokushisa elilinganisiwe newashi lesistimu.
●Xhumana nekhompyutha ngesixhumi esibonakalayo se-RS232 esijwayelekile.
Ukunikezwa kwamandla kagesi | I-AC220V±10%, 50Hz |
I-voltage yokuphumayo | 0~80kV(Kukhethiwe) |
Amandla | 1.5kVA |
Amandla | 200W |
Isivinini sokukhuphuka kwe-voltage | 2.0~3.5kV/s (Iyalungiseka) |
Amandla kagesiukulinganisaukunemba | ±3% |
Ukuhlanekezela kwe-Waveform | ≤3% |
Isikhawu sokuthuthukisa | 5min (Iyalungiseka) |
Isikhathi sokuma | 15min (Iyalungiseka) |
Izikhathi zokuthuthukisa | 1~6 (Kukhethwa) |
Isebenzaimvelo | Tizinga lokushisa: 0℃-45°C |
Hubuncane:Max.umswakama ohlobene75% | |
Ubukhulu | 465x385x425mm |
① Iphrinta eshisayo--iphrinta imiphumela yokuhlolwa;
② LCD--ibonisa imenyu, ngokushesha kanye nemiphumela yokuhlolwa;
③ Okhiye bokusebenza:
Cindezela inkinobho ethi "◄" ukuze ukhuphule inani lokusetha;
Cindezela inkinobho ethi "►" ukuze unciphise inani lokusetha;
Khetha--ukukhetha imisebenzi (into ekhethiwe isesibukweni esibekiwe);
Qinisekisa--ukwenza imisebenzi;
Emuva--yokuphuma ku-interface yokusebenza;
④ Ukushintsha amandla kanye nenkomba
1. Ukulungiselela ngaphambi kokuhlolwa
1.1 Xhuma itheminali yaphansi (ngakwesokudla sempahla) ocingweni lwaphansi ngokuqinile ngaphambi kokusebenzisa okokusebenza.
1.2 Khipha isampula likawoyela ngokwezinga elifanele.Lungisa ibanga le-electrode ngaphakathi kwenkomishi kawoyela njengegeji ngayinye evamile.Hlanza inkomishi ngokwezidingo ezifanele.Thela isampula yamafutha enkomishini bese uvala ikepisi.
1.3 Ukushintsha ugesi we-AC220V ngemva kokuqinisekiswa kwezinto ezingenhla, zilungele ukuhlolwa.
2. Ukuhlola
2.1 Cindezela inkinobho yamandla bese ungena kusixhumi esibonakalayo esilandelayo:
2.2 Ukusetha Ipharamitha Yesistimu
Cindezela inkinobho ethi "Qinisekisa" bese ufaka esibonakalayo esilandelayo:
Isilungiselelo esikhuthazayo: abasebenzisi bangakhetha ngokuya ngesidingo sangempela.
Cindezela inkinobho ethi "Emuva" ukuze uphume kulesi sixhumanisi ngemva kokwenziwa kokulungiselelwa.
2.3 Ukuhlola
Cindezela inkinobho ethi "Khetha" ukuze ukhethe imenyu ethi "Qala Ukuhlola" bese ucindezela ukhiye "Qinisekisa" ukuze ungene kusixhumi esibonakalayo esilandelayo:
Ukuze uqhubeke nokuhlola okulandelayo ngokushesha nje lapho uhlolo lokuqala seluqediwe kuze kuqedwe isethi yokwandisa imvamisa.Ekugcineni, umphumela uyavezwa futhi uphrintwe ngale ndlela elandelayo:
2.4 Ukubuka nokuphrinta idatha:
Cindezela inkinobho ethi "Khetha" ukuze ukhethe imenyu "Yokubuka Nokuphrinta Idatha" bese ucindezela ukhiye "Qinisekisa" ukuze ungene kusixhumi esibonakalayo esilandelayo:
Khetha "Ikhasi Eliphezulu" noma "Ikhasi Phansi" bese ukhetha amarekhodi azophrintwa bese ukhetha "Phrinta".
●Ukukhethwa kwesampula kawoyela kanye nokubekwa kwebanga le-electrode kuzohlangabezana namazinga afanelekile kazwelonke nawemboni.
●Abaqhubi noma abanye abasebenzi banqatshelwe ngokuphelele ukuthinta igobolondo ngemva kokuba amandla evuliwe ukugwema izingozi.
●Amandla azonqanyulwa ngokushesha uma kukhona isigameko esingavamile esitholakala ngesikhathi sokuhlinzwa.
●Lokhu okokusebenza akufanele kuvezwe endaweni enomswakama.
●Gcina inkomishi kawoyela nama-electrode kuhlanzekile.Gcwalisa i-inkomishi enamafutha e-transformer entsha ukuze ivikeleke uma ingenzi lutho.Hlola ibanga le-electrode futhi uhlole ukucinana phakathi kwethiphu ye-electrode kanye nentambo yesikulufu sebha ye-electrode ngaphambi kokuthi inkomishi iphinde isetshenziswe.
1. Indlela Yokuhlanza Indebe Yamafutha
1.1 Sula izindawo zama-electrode kanye nezinsimbi kaningi ngendwangu kasilika ehlanzekile.
1.2 Lungisa ibanga le-electrode ngegeji evamile
1.3 Sebenzisa i-petroleum ether (ezinye izinyibilikisi zemvelo azivunyelwe) ukuhlanza izikhathi ezintathu.Isikhathi ngasinye kuzolandela izinqubo ezilandelayo:
① Thela i-petroleum ether enkomishini kawoyela ize igcwale ngo-1/4~1/3.
② Mboza unqenqema lwenkomishi ngocezu lwengilazi ehlanzwe nge-petroleum ether.Gubha inkomishi ngokulinganayo umzuzu owodwa ngamandla athile.
③ Thela i-petroleum ether bese womisa inkomishi nge- blower imizuzu engu-2~3.
1.4 Sebenzisa isampula likawoyela elizohlolwa ukuze uhlanze inkomishi izikhathi ezingu-1–3.
① Thela i-petroleum ether enkomishini kawoyela ize igcwale ngo-1/4~1/3.
② Mboza unqenqema lwenkomishi ngocezu lwengilazi ehlanzwe nge-petroleum ether.Gubha inkomishi ngokulinganayo umzuzu owodwa ngamandla athile.
③ Thela isampula yamafutha engakwesokunxele bese kuqala ukuhlola.
2. Indlela Yokuhlanza Induku Eshukumisayo
2.1 Sula induku enyakazayo futhi ngendwangu kasilika ehlanzekile kuze kube yilapho izinhlayiya ezincane zingatholakali ezindaweni zazo.Akuvunyelwe ukuthinta izindawo ezingaphezulu ngezandla.
2.2 Sebenzisa izindlawu ukuze ubambe induku;uwafake ku-petroleum ether bese uwageza.
2.3 Sebenzisa ama-forceps ukuze ubambe induku futhi uyomise nge- blower.
2.4 Sebenzisa izindlawu ukuze ubambe induku;uwafake esitsheni samafutha bese uwageza.
3. I-Oil Cup Storage
Indlela yoku-1 Gcwalisa inkomishi ngamafutha amahle okuvikela ngemuva kokuphela kokuhlolwa bese uyibeka izinzile.
Indlela yesi-2 Hlanza bese womisa inkomishi ngaphansi kwalezi zinqubo ezingenhla bese uyifaka endaweni yokomisa ivacuum.
Qaphela: Inkomishi yamafutha kanye nenduku eshukumisayo izohlanzwa ngaphansi kwezinqubo ezingenhla ngemva kokuhlolwa kokuqala nokuhlolwa ngamafutha ampofu.
4. Common Fault Clearances
4.1 Amandla acishiwe, asikho isibonisi esikrinini
① Hlola ukuthi ipulaki lamandla lifakwe liqinile noma cha.
② Hlola ukuthi i-fuse engaphakathi kwe-outlet yamandla isesimweni esihle noma cha.
③ Hlola ugesi wesokhethi.
4.2 Inkomishi kawoyela ngaphandle kokuphazamiseka
① Hlola ukufakwa kwezixhumi ebhodini lesekethe.
② Hlola ukuthintwa kweswishi enevoltage ephezulu kukhava yekesi.
③ Hlola ukusebenza koxhumana nabo bamandla kagesi aphezulu.
④ Hlola ukunqamuka komugqa wamandla kagesi aphezulu.
4.3 Ukungafani kokuboniswa akwanele
Lungisa i-potentiometer ebhodini lesifunda.
4.4 Ukwehluleka kwephrinta
① Hlola ipulaki yamandla yephrinta.
② Hlola ukuxhunywa kolayini wedatha yephrinta.