I-GDYB-S3 ISigaba Sesithathu Sokuhlola Imitha Yamandla (izikhundla ezi-3)
Ukunemba kwesistimu: 0.1%;
Ipharamitha yomthombo wezigaba ezintathu:
a.I-voltage yokuphumayo:
3*57.7VY 3*100V ∆ ;
3*220V Y 3*380V ∆;
Ububanzi obulungisekayo 0~120% obulungiswa ngokuqhubekayo
Ukulungiswa kokulunga: okungcono kuno-0.01%
b.Okukhiphayo okwamanje: 25mA, 0.1A, 1A, 5A, 25A, 100A;
Ububanzi obulungisekayo 0~120% obulungiswa ngokuqhubekayo
Ukulungiswa kokulunga: okungcono kuno-0.01%
Okukhiphayo okuncane okwamanje: 1mA (ukunemba: <5%)
c.Amandla okukhipha: (i-voltage circuit) 3 * 50VA
(isifunda samanje) 3 * 150VA
d.I-voltage ephumayo, yamanje, ukuzinza kwamandla: ≤0.02%/120S
e.Idigri yokuhlanekezela ye-voltage ephumayo kanye ne-waveform yamanje: ≤0.5%
f.I-Symmetry yama-engeli ezigaba ezintathu: phakathi kuka-120°±0.3°
g.Ibanga lemvamisa: 45Hz~ 65Hz, isinyathelo: 0.01Hz
h.Ibanga le-engeli yesigaba: -180°~ 180°, isinyathelo: 0.01°
i.Ibonisa ukunemba kwemitha: I-Voltmeter, i-ammeter: ±0.01%
Imitha yesigaba: ±0.01°
Imitha yemvamisa: ±0.01Hz
Ubuncane bomphumela wamanje: 1mA;ukunemba: ≤5%;
Umthamo wokulayisha okukhiphayo (amamitha angama-20)
Okukhipha amandla kagesi: 800VA;
Okukhiphayo kwamanje: 2500VA;
I-Voltage nokulungiswa kwamanje
Ibanga elingalungiseka: (0%~120%) x ububanzi bokulinganisa;
Ukucwiliswa okulungisekayo: 10%, 1%, 0.1%, 0.01%;
Ukulungiswa kwesigaba:
Ububanzi obulungisekayo: 0°~360°;
Ubucwebe obulungisekayo: 10°, 1°, 0.1°, 0.01°;
Ukulungiswa kwemvamisa:
Ububanzi obulungisekayo: 45Hz ~ 65Hz;
Ubunono obulungisekayo: 1Hz, 0.1Hz, 0.01Hz;
Ukuqina kwamandla okukhiphayo: ≤0.05%/120S;
I-voltage yokuphumayo nokuhlanekezelwa kwamanje:
Amandla kagesi: ≤0.5%;
Okwamanje: ≤0.5%;
Izilungiselelo ze-Harmonic:
Inombolo ye-Harmonic: izikhathi ezingu-2 ~ 21;
Okuqukethwe kwe-Harmonic: izikhathi zamanje ze-2 ~ 21 ≤40% (ezihlobene negagasi eliyisisekelo);
i-voltage 2 ~ izikhathi ezingu-21 ≤40% (ngokuhlobene namagagasi ayisisekelo);
Isigaba se-Harmonic: 0 ° ~ 359.99 °;
Ithuluzi lokuqapha:
Ukunemba kokubonisa imitha ejwayelekile: Amandla kagesi: 0.01%;
Okwamanje: 0.01%;
Amandla: 0.01%;
Isigaba: 0.01;
Imvamisa: 0.01;
Insulated 1:1 ipharamitha ye-transformer yamanje:
Ukunemba: 0.01%;
Amandla okukhipha: 60VA;
Ibanga lamanje lokulinganisa: 1mA~120A (1:1);
Inkomba yokusebenza eyinhloko
a) Ukunemba: 0.01% (COS∮=0.5C…1…0.5L);
b) Imvamisa elinganiselwe: 50Hz/60Hz±5%;
c) Okuyisisekelo samanje: 3 × 0.001-120A;
d) Yamanje yesibili: 3 × 0.001-120A;
e) Umthwalo wokuphumayo (isigaba ngasinye):
0.5W (ububanzi bokukhiphayo: 0.001-1A);
0.7V ihlukaniswe ivelu yamanje (ububanzi bokuphumayo: 1-120A);
f) I-voltage yomthwalo wokuphumayo: Ubukhulu.0.5V (output current Max. 120A);
Ubukhulu be-transformer yamanje: 265 x 200 x 240 mm (L x W x H);
Isisindo: 12kg;
Ubukhulu:
Umthombo wokulawula: 610 x 800 x 1920mm (L x W x H);
Imitha yokubeka: 2350 x 750 x 1850mm (L x W x H);