I-GDYM-3F Portable Multi-functional Energy Meter Calibrator
●Ngokuvumelanayo ukubala yonke imingcele nokuboniswa kusixhumi esibonakalayo kufaka phakathi igagasi eligcwele, igagasi eliyisisekelo negagasi elivumelanayo njengoba ngesikhathi esifanayo bonisa u-U, I, P, Q, VA, i-angle yesigaba mayelana negagasi eligcwele kanye namagagasi ayisisekelo kanye nokuhlangana kwe-harmonic kwe-I, U, P, Q, VA kanye nephesenti lakho, okuqukethwe kwe-U, ukulandelana okuhle kwe-I, ukulandelana okunegethivu, ukulandelana kweziro nokuvumelana kwe-PH+, PH-, PH+/Pf%, PH-/Pf%, igagasi eligcwele PF, igagasi eliyisisekelo PF, igagasi le-harmonic PF.
●Indlela yokuhlola ihlanganisa izinhlobo ezingu-7 zegagasi eligcwele elisebenzayo, amandla asebenzayo agcwele amagagasi, igagasi eligcwele elibonakalayo, igagasi eliyisisekelo elisebenzayo, amandla ayisisekelo asebenzayo egagasi, igagasi eliyisisekelo elibonakalayo kanye ne-harmonic wave, engahlola ukunemba kwamandla asebenzayo emitha yegagasi eligcwele, okuyisisekelo. imitha, i-harmonic wave meter kanye nokunemba kwamandla asebenzayo kwe-wave wave ephelele kanye nemitha ye-harmonic wave, nakho kungabheka ukunemba okusobala kwemitha yamandla kagesi.
●Uma uthanda i-harmonic eyodwa noma ngaphezulu ukubala inhlanganisela ye-harmonic ye-U, I, P, Q, VA kanye nephesenti layo kanye namandla asebenzayo we-harmonic wave wamandla kagesi.
●Ngesikhathi esifanayo ingxenye yokuhlaziya eyodwa kanye nenani lokulinganisa lesikhathi ngasinye i-U, I, P, Q, VA yezigaba ezintathu zamagagasi e-harmonic futhi ibonise izikhathi zayo ezingu-51 +PH/-PH kanye nephesenti legagasi eliyisisekelo ukuze kunqunywe ithonya le-harmonic enhle nebi imitha.
●Amandla asebenzayo wamagagasi agcwele iyisamba se-algebraic samandla asebenzayo wegagasi eliyisisekelo namagagasi avumelanayo e-harmonic ngayinye.Ukuze ugweme ukushintsha kwesigaba se-90 degree esikhathini esidlule kanye nasesimeni sephutha lokuhlola amandla asebenzayo we-harmonic wave.Singakala umkhiqizo ngemitha yamandla kagesi engatholwa bese sikhetha inombolo yegagasi ngalinye elihambisanayo.
●Ngesikhathi esifanayo hlola ukushaya kwenhliziyo okubili okutholiwe, futhi imodi engaguquki, yokusebenza kanye nenani lama-pulses atholiwe ngesikhathi esisodwa kungase kwehluke ngokuphelele.
●Ukulayisha ngokweqile izikhathi ezingu-2.4, wonke amaphuzu okulinganisa angaqinisekisa ukuthi ukunemba kuka-0.05% futhi aqinisekisa inkomba yokulinganisa ye-CT yesibili elinganiselwe yamanje uma ingu-1A.
●Izikhathi ezingu-500 ububanzi bamanje buqinisekisa inkomba yokulinganisa, ibanga lokubonisa lezikhathi ezingu-1000, ubuncane obungu-1mA obuqalayo okwamanje.Ngesikhathi esifanayo sokunikezwa kwamandla kagesi aphezulu, idivayisi ye-high-voltage iyalinganiswa, futhi i-wiring ingabonakala ngokungabi nomthwalo.
●Umsebenzi wokuqongelela amandla kagesi ungakala imitha yamandla kagesi engenawo ukushaya kwenhliziyo.
●Bonisa i-vector diagram, ingakwazi ukubona izintambo ezintathu zezintambo ezingalungile ezingu-48 kanye nezintambo ezine ezinezintambo ezingalungile ezingu-96 kanye nezintambo eziphambanayo zesibili ze-CT.
●I-Adopt 7 intshi 16:9 isikrini sombala wezinga lemboni nobubanzi be-TFT, ipharamitha ngayinye yokusebenza yethuluzi iwukuboniswa okuqondile nokucacile esikrinini ngesikhathi esisodwa ngaphandle kokuvula isikrini , kuvumela abasebenzisi ukuthi basebenzise ngokubuka nje.
●Insimbi ingalingisa ukuhlukahluka kwezintambo zesigcawu futhi ithole imiphumela ehambisanayo ye-wiring ne-coefficient yokulungisa ehhovisi ngaphandle kokufinyelela ku-voltage noma isignali yamanje.Lo msebenzi ungasetshenziswa njengethuluzi lokuhlola umugqa wokuqeqesha ukuthuthukisa amakhono okuhlola umugqa wabasebenzi basensimini.
●Insimbi yangaphakathi ingakhiqiza isignali yangempela yamanje engu-0.05-5A futhi yengezwe kusigxivizo esingu-5A ukuhlola iphutha le-clamp meter.Ukucabangela i-5A clamp inenkinga yokushintsha iphutha ngemuva kokusetshenziswa isikhathi eside.Nika amandla ukwazi isimo sokuvala mayelana nemitha yokubopha nganoma yisiphi isikhathi, khumbuza abasebenzi abafanelekile ngesikhathi ukuthi bahlanze umhlathi wensimbi.Lapho ushintsha i-clamp engu-5A, iphutha lingalungiswa libe ngu-0.2.
●Faka ngokuzenzakalelayo inombolo yemitha yamandla (umsebenzi ozikhethela wona)
Sekela isithwebuli sebhakhodi futhi ufake ngokuzenzakalelayo ikhodi yemitha yamandla.
●Sebenzisa i-485 noma infrared photoelectric head ukuze ufunde ithiphu lemitha lamandla emisebenzi eminingi, isiqongo, isicaba, isigodi, isisekelo esisebenzayo nesisebenzayo, funda njalo kanye nesikhathi.Sekela imithetho kazwelonke ejwayelekile ye-DL645, i-DL645-1997 kanye ne-DL645-2007.(umsebenzi ozikhethela wona)
●Idiski ye-electronic eyakhelwe ngaphakathi ye-2G.
●Ingahlonywa ngephrinta yenkambu yamayintshi angu-3, esebenza ngebhethri, futhi iguqule ngokuqondile isithombe sesikrini sombala sibe mnyama nokumhlophe uma siphrinta.(ongakukhetha)
●Ukusebenzisa imitha yokubopha (5A, 100A, 500A, 1000A) ngokuqondile ukukala isilinganiso nephutha lesilinganiso se-voltage ephansi ye-transformer yamanje.(100A, 500A, 1000A isigxivizo uyazikhethela)
●Bonisa i-voltage ne-waveform yamanje yezigaba ezintathu A, B, C.
●Ububanzi bokufaka amandla okusebenza ngu-55V-600V.Isokhethi yezimboni ezijwayelekile zezigaba ezintathu kagesi ngu-600V.Amaswishi emboni ashintsha ukuxhumana kukagesi wedolobha kanye ne-voltage yezigaba ezintathu njengamandla omsebenzi.
●Yonke imingcele edinga ukufakwa esizeni kanye nedatha elandiwe ingachazwa ngokukhululekile ngumsebenzisi.Ingakwazi ukulanda idatha yabasebenzisi abangu-5000.
●Isikrini samanje singagcinwa kudiski ye-U.
●Ngomsebenzi wokulinganisa izinga lokushisa nomswakama (isidingo sokufaka inzwa yokushisa neyomswakama).
●Nge-USB interface futhi ixhumeka ku-USB flash drive ukwakha isikhala esikhulu sokukopisha idatha.
●Isikrini sokuthinta bese ucindezela ukusebenza kokhiye, ikhibhodi isebenzisa ukuhlelwa kwekholomu yekhibhodi yekhompyutha.Ikhibhodi ye-PC ene-USB interface iyatholakala.
●Isikhathi ngomshini we-PC ngesixhumi esibonakalayo se-USB.
●Ikhompuyutha yethebhulethi ingakwazi ukulawula ithuluzi ngokuqondile ngesixhumi esibonakalayo esisebenzayo njengethuluzi.
●Isokhethi yokulinganisa i-voltage kanye neswishi yokushintsha amandla ebangeni lemboni ingasetshenziswa ngaphansi kuka-600V (Iswishi yamandla esetshenziswayo evamile ingu-250V, esetshenziswa ebangeni elingaphezu kuka-380V, elinezingozi zokuphepha ezingaba khona).
●Idizayini yokuphepha ephezulu.Isokhethi kagesi, isokhethi yokukala i-voltage, okokufaka kwamanje, okokufaka kwe-clamp yamanje kuhlukanisiwe, kanti i-standard voltage ingu-4kV.
●Ukuvikelwa kokufakwa kwe-pulse yamandla kagesi.Uma ifaka i-voltage ephezulu izoqala ngokuzenzakalelayo ukuvikela futhi ibonise umlayezo osheshayo ensimbini, lapho iphutha selilungisiwe, chofoza isixhumi esibonakalayo ukuze usethe kabusha ukuvikela.
●I-5A clamp inomsebenzi wokuzilungisa ngokwayo: isignali yamanje ye-5A ikhiqizwa ngaphakathi ithuluzi elingezwe ku-clamp engu-5A ukukala iphutha lemitha yokubopha.
●Kusetshenziswa ama-chips e-ARM angu-5 amandla aphansi, wonke amandla okubala makhulu kuno-1GIPS, futhi yonke imingcele ye-harmonic engu-51 kanye negagasi eligcwele libalwa ngesikhathi esifanayo.
●Ukusebenzisa uhlelo lokusebenza lwe-Windows CE oluvuthiwe futhi lwangempela, umsebenzi walo uzinzile futhi uthembekile.
Amandla kagesi | Ibanga lephutha lesiqinisekiso | 25-600V | |||
Iqala inani lokulinganisa | 0.5V | ||||
Inani elilinganiselwe | 440V | ||||
Ukunemba kwephutha | ±0.05% | ||||
Okwamanje | Isikhumulo | IkaliweIe | 5A | ||
Ibanga | 0.025A-12A | ||||
I-Clamp | IkaliweIe | 5A, 100A, 500A,1000A(uyazikhethela) | |||
Ibanga | 0.5%Ie—2.4Ie | ||||
Iqala yamanje | 0.02%Ie | ||||
Ibanga lefrikhwensi yokufaka | 45-65Hz | ||||
Iphutha lokulinganisa imvamisa | ±0.01Hz | ||||
Ibanga lokulinganisa i-engeli yesigaba | -180°-180° | ||||
Ubukhulu.Iphutha le-engeli yesigaba | ±0.1° | ||||
Amaphesenti e-Harmonic, amapharamitha kagesi e-harmonic, nezikhathi zokulinganisa amandla kagesi e-harmonic | 2-51izikhathi | ||||
I-Pulse njalo (Qaphela:Ieilinganiselwe okwamanje) | Okukhipha imvamisa ephansi(3 amagiya) | 5A | 3600P, 36000P, 360000P/kWh(kvarh, kVAh) | ||
abanye | 3600P, 36000P, 360000P *(5/Ie)P/kWh (kvar, kVAh) | ||||
Imvamisa ephezulu yangaphakathi | 5A | 1.8x109I-P/kWh | |||
Amanye amagiya amanje | 1.8x109*(5/Ie)P/kWh | ||||
Imvamisa yokufaka ishayela ephezulu | 50kHz | ||||
5AIsikhumulo | Igagasi eliyisisekelo legagasi eligcwelekWh, P, U, I | ±0.05% | |||
Igagasi eliyisisekelo legagasi eligcwele Kvarh, Q, VA, VAh | ±0.2% | ||||
I-HarmonicigagasiU, I, P, Q, S | ±0.2%H(H=1+0.01k K=2-51) | ||||
5Aqinisa | Igagasi eliyisisekelo legagasi eligcwelekWh, P, U, I | ±0.2% | |||
Igagasi eliyisisekelo legagasi eligcwele Kvarh, Q, VA, VAh | ±0.2% | ||||
I-HarmonicigagasiU, I, P, Q, S | ±0.5% H(H=1+0.01k K=2-51) | ||||
Ukuzihlola kwamanje okuphumayo kwe-cyclic | 0.05A, 0.1A, 0.25A, 0.5A, 1A, 2.5A, 5A | ||||
Uqobo lwakhoiyahlolaiphutha (umehluko womugqa kanye ne-angular) | ±0.1% | ||||
I-clamp yamanje enkulu | Igagasi eliyisisekelo legagasi eligcwelekWh, P, U, I | ±0.5% | |||
Igagasi eliyisisekelo legagasi eligcwele Kvarh, Q, VA, VAh | ±0.5% | ||||
I-HarmonicigagasiU, I, P, Q, S | ±1.0% H(H=1+0.01k K=2-51) | ||||
Abanye | Isilinganiso sokukala | ±0.5% | |||
Ithonya lamandla kagesi | < ±0.01% | ||||
Imvamisaithonya | < ±0.01% | ||||
Izinga lokushisaithonya | < ±0.005%/℃ | ||||
Iphutha lokushintsha lamahora angu-24 | < ±0.01% | ||||
I-Harmonicithonya | < ±0.01% | ||||
Ibhethri elishajekayo | 11.1V 3000mAh | ||||
Ukumelana ne-voltage (Isilinganiso samanje se-Voltge kanye nokunikezwa kwamandla kanye namatheminali ane-voltage ephansi) | 4.0 kV | ||||
Ukuqhuma Okudlula Kagesi Okusheshayo | 4.0 kV | ||||
Surge | 4.0kV(Imodi evamile),2.0kV(Imodi ehlukile) | ||||
Isikhathi sokulinda ngemuva kokuqala | <3umzuzu | ||||
Izinga lokushisa le-ambient | -25℃-+45℃ | ||||
Umswakama ohlobene | 40%-95% | ||||
Ukunikezwa kwamandla okusebenzayo | I-AC 55V - 600V | ||||
Ukusetshenziswa kwamandla | < 9W | ||||
Ubukhulu | 275mm * 196mm * 78mm | ||||
Isisindo | < 1.5 kg |
1--Thinta isikrini
2--Ibha yenqubo
3--Susa ukhiye
4--Phuma(emuva) ukhiye
5--Ikhibhodi
6--Shintsha ukhiye
7--Ukhiye wesikhala
8--Ukhiye wokuzulazula/KULUNGILE
9--Isokhethi yamandla
10--Ukushintsha kwamandla
11--Isokhethi yezigaba ezintathu zokufakwa kwamandla kagesi (ABC)
12--Imbobo ye-USB
13--RS232 imbobo yokulungisa isixhumi esibonakalayo
14--Isokhethi yokushaja
15--Ukushintsha ibhethri
17--Isibambi sezintambo eziyisikhombisa se-A phase current clamp
18--Isibambi sezintambo eziyisikhombisa sesigaba B se-clamp yamanje
19--Isibambi sezintambo esiyinhloko esiyisikhombisa se-clamp yamanje yesigaba C
20--Isixhumi esibonakalayo se-Pulse
21--Isigaba samanje sokufakwayo nesokhethi yokuphumayo
22--B isigaba sokufakwayo samanje nesokhethi yokuphumayo
23--C isigaba sokufakwayo samanje nesokhethi yokuphumayo
24--Ibhande elilengayo
Isilinganisi senkundla semitha yesigaba sesithathu se-watt | 1 isethi |
Iwashi le-clamp engu-100A | 3 izingcezu |
Indandatho yokulungisa i-clamp | 1 isiqephu |
Ikhanda likagesi wesithombe df307 | 1 isiqephu |
Idiski ye-USB (isoftware yokuphatha) | 1 isiqephu |
I-3 phase voltage line high voltage (kufaka phakathi i-clamp nephini) | 1 isiqephu |
Umugqa wamanje wesigaba esingu-3 (okuhlanganisa i-clamp kanye nokufaka) | 1 isiqephu |
Umugqa we-pulse wesigaba esi-3 1A | 1 isiqephu |
I-adaptha | 1 isiqephu |
Intambo yamandla | 1 isiqephu |
Imichilo | 1 isiqephu |
Ibhokisi lezinsimbi | 1 isiqephu |
Imanuwali yomkhiqizo | 1 ikhophi |
Isitifiketi somkhiqizo | 1 ikhophi |
Uhlu lwamapaki embonini | 1 ikhophi |
Umbiko wokuhlolwa kwefekthri | 1 ikhophi |