I-HV HIPOT GDC-9560B Power system Insulation Oil Gas Chromatography Analyzer
Ukufakwa nokususwa kwekholomu ye-chromatographic:
1. Ukufakwa nokususwa kwekholomu ye-chromatographic kufanele kwenziwe ekamelweni lokushisa.
2. Imibhoshongo epakishiwe inezimpawu ze-ferrule kanye ne-gasket seal.Kunezinhlobo ezintathu zama-ferrules: izinsimbi zensimbi, izinsimbi zepulasitiki nama-graphite ferrules, okungelula ukuthi ziqiniswe kakhulu uma zifakwa.Izigxivizo zohlobo lwe-gasket zidinga ama-gasket amasha isikhathi ngasinye lapho kufakwa ikholomu.
3. Ukuthi iziphetho ezimbili zekholomu yechromatographic zixhunywe ngoboya bengilazi yini.Vimbela uvolo wengilazi nokupakishwa ukuthi kungapheshulwa kumtshina womoya yigesi yenkampani.
4. Ubude bokufakwa nokufakwa kwekholomu ye-capillary kuncike encwadini yeziqondiso zethuluzi.Isakhiwo samagumbi ahlukene we-chromatographic evaporation sihlukile, futhi ubude bokufaka buhlukile.Kufanele kuqashelwe ukuthi uma ikholomu ye-capillary isetshenziswa ngokugeleza okungahlukanisiwe, ukuxhumana phakathi kwegumbi lokuhwamuka kanye nekholomu epakishiwe akumele kube nama-probe amaningi axhunywe egumbini lokuhwamuka, futhi ikholomu ye-capillary kufanele ibe ngaphezu kancane kwe-cap.
Umthelela wesilinganiso se-hydrogen emoyeni kumtshina we-FID:
Ku-chromatograph yegesi, isilinganiso se-hydrogen nomoya kufanele sibe ngu-1:10.Uma isilinganiso siphezulu kakhulu, ukuzwela komtshina we-hydrogen flame kuzokwehla kakhulu.Hlola amazinga e-hydrogen nomoya wokugeleza.Uma igesi ku-hydrogen nomoya inganele, izokwenza "i-bang" yokushisa, bese icisha umlilo, ngokuvamile lapho ukhanyisa isitofu sikagesi, siyacishwa, bese siphinde sikhanyise, bese i-hydrogen ayanele.
Ungavimba kanjani inaliti yomjovo ukuthi ingagobi:
Abaqalayo abaningi ku-chromatography bavame ukugoba inaliti nesiqu ngoba:
1. Imbobo yomjovo ye-chromatograph yegesi isongwe ngokuqinile, futhi igoqwe ngokuqinile kakhulu ezingeni lokushisa legumbi.Lapho izinga lokushisa lekamelo le-vaporization liqhubeka nokukhuphuka, i-silicone gasket izokhula futhi iqinise.Ngalesi sikhathi, kunzima ukungena ngesirinji.
2. Inaliti ibhajwe engxenyeni yensimbi yembobo yomjovo uma indawo ingelula ukuyithola.
3. Induku yesirinji igobile ngoba amandla anamandla kakhulu lapho ujova isampula.Kukhona i-sampler rack yokungenisa nge-chromatographic.Akunakwenzeka ukugoba induku yesirinji evamile lapho usebenzisa i-sampler rack ukujova amasampula.
4. Njengoba udonga lwangaphakathi lwesirinji lungcolile, phusha umphini yenaliti ukuze ugobe ngesikhathi sokujova.
5. I-chromatograph yegesi kufanele izinzile ngesikhathi sokujova.Uma ujahile uzogoba umjovo.Uma ujwayele lo msebenzi, uzoqedwa ngokushesha.
Isikhathi sokuthumela: Jun-07-2022